2-5
connection allows for interfacing to an externally located 24 bits, 800 x 600, SVGA
monitor, which should be a medical grade monitor.
8. USB
port
9. Network
connection
One multiplex connector for network and software online upgrade.
10. Nurse call connection
Connects to the hospital’s calling system and outputs nurse call signals when an alarm
occurs.
11. Inspiratory dust filter
12. Fan filter (filter at air intake vent)
13. Equipotential stud / lug
Provides a ground point. Eliminates the ground potential difference between different
devices to ensure safety.
2.2.3 Air Compressor
The air compressor has standby function. In the standby mode, the compressor starts to
deliver compressed air to the ventilator automatically if the hospital central pipeline gas
supply fails. The compressor stops delivering compressed air automatically when the central
pipeline gas supply returns to normal.
1. Power
indicator
The power indicator is lit when the compressor is connected to power supply and the
power switch is turned on.
2. Status
indicator
The status indicator is lit when the central pipeline gas supply is applied.
1 2 3
4
5
6
7 8 9 10
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...