6-19
In DuoLevel under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Phigh
]:
High
pressure
3. [
Thigh
]:
Time
of
high
pressure
4. [
Plow
]:
Low
pressure
5. [
Tlow
]:
Time
of
low
pressure
6. [
Tslope
]:
Time
of
pressure
rising
7.
[
Exp
%
]:
Expiration
trigger
level
8. [
△
Psupp
]:
Pressure support level
9. [
Δ
Papnea
]:
Pressure
of
apnea
ventilation
10. [
fapnea
]:
Frequency
of
apnea
ventilation
11. [
F-Trig
] or [
P-Trig
]: Inspiration
trigger
level
6.7.10 APRV
APRV is airway pressure release ventilation. It can be seen as periodical, short period airway
pressure release in CPAP mode.
The following figure shows typical waveform in APRV.
Paw
Time
Phigh
Tslope
Thigh
Plow
Tlow
In APRV under non-NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen concentration
2. [
Phigh
]:
High pressure
3. [
Thigh
]:
Time of high pressure
4. [
Plow
]: Low
pressure
5. [
Tlow
]:
Time of low pressure
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...