6-11
In P-A/C under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Pinsp
]:
Pressure
control
level
of
inspiration
3. [
Tinsp
] or [
I:E
]:
Time of inspiration or ratio between the inspiratory and
expiratory time
4. [
f
]:
Breathing
frequency
5. [
Tslope
]:
Time
of
pressure
rising
6 [
PEEP
]:
Positive
end-expiratory
pressure
7. [
Assist
]:
Assisted
trigger
8. [
F-Trig
] or [
P-Trig
]: Inspiration
trigger
level
6.7.6 CPAP/PSV
PSV, pressure support ventilation, is a pressure mode. The system delivers a PSV when it
detects that patient inspiratory effort reaches the preset inspiration trigger level. Time of
pressure rising and pressure support level are set by the user. At the beginning of inspiratory
phase, patient airway pressure rises to the preset pressure level within the preset Time of
pressure rising and is held at this pressure level until patient inspiratory flow is detected to
reach the expiration trigger level. In PSV, when the airway pressure is held at the preset
pressure level, delivered gas flow changes with the resistance and compliance of the patient’s
lungs.
Paw
Flow
Time
Apnea time
△
Papnea
Apnea ventilation
Apnea vent cycle
I:E=1:2
Tslope
△
Psupp
PSV
Exp. trigger
Insp. trigger
Insp. trigger
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...