D-3
personnel.
RT Clock Need
Reset
L
There is button cell available in the system but the clock is
reset in case of power failure. Contact your service
personnel.
RT Clock Not
Exist
H
There is no button cell available in the system or the
battery has no capacity.
The clock chip is faulty. Contact your service personnel.
IP Address
Conflict
L
This message is displayed when there are conflicts
between IP address settings and disappears until the
conflicts are resolved. Set IP address again.
Loading Default
Config. Failed
L
Loading configuration is failed.
Reload the configuration. If the problem persists, contact
your service personnel.
Restoring Last
Config. Failed
L
Restoring configuration is failed.
Restart the ventilator or reload the configuration. If the
problem persists, contact your service personnel.
SD Storage Card
Error
L
Storage card is not available or is failed.
Replace the storage card.
Insp. Hold
Interrupted
L
This alarm is triggered when the Insp. Hold key is pressed
and held longer than the time limit. The alarm is cancelled
when the Insp. Hold key is released.
Exp. Hold
Interrupted
L
This alarm is triggered when the Exp. Hold key is pressed
and held longer than the time limit. The alarm is cancelled
when the Exp. Hold key is released.
Keyboard Comm
Stop
H
Keyboard malfunction. Contact your service personnel.
Keyboard Selftest
Error
H
Keyboard selftest error. Contact your service personnel.
Ventilator Reset
Error
H
One of ventilator modules is powered on abnormally. Push
the Alarm Reset key to cancel this alarm.
When this alarm is triggered, press [
Alarm Reset
] key and
the alarm disappears.
Key Error
M This alarm is triggered when the hard key or rotary
encoder is pressed and held for more than 35s.
Memory Error
H
The memory has an error. Restart the ventilator. If the
problem persists, contact your service personnel.
PCON2 Selftest
Error
H
The key control board has an error during power-on.
Restart the ventilator. If the problem persists, contact your
service personnel.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...