B.8 Alarms
B.8.1 Settable Alarms
Alarm settings
Parameter
Setting range
Automatic threshold Notes
TV High limit 110 to 4000 mL, OFF(Adult)
25 to 4000 mL, OFF(Pediatric)
1.5 × TVe average
value
High limit
is greater
than low
limit.
MV High limit Pediatric: 0.2 to 60.0 L/min (non-NIV)
Adult: 0.2 to 100.0 L/min (non-NIV)
Pediatric: 0.1 to 60.0 L/min (NIV)
Adult: 0.1 to 100.0 L/min (NIV)
1.5 × MV monitored
value
Low limit
Pediatric: 0.1 to 30.0 L/min (non-NIV)
Adult: 0.1 to 50.0 L/min (non-NIV)
Pediatric: 0.1 to 30.0 L/min, OFF (NIV)
Adult: 0.1 to 50.0 L/min, OFF (NIV)
0.5 × MV monitored
value
Paw High limit 10 to 105 cmH
2
O Average
peak
p10cmH
2
O
and 35cmH
2
O,
whichever is greater
ftot
High limit 1 to 160, OFF
1.4 × ftot monitored
value
Tapnea
5 to 60s
15
/
B.8.2 Auto Alarms
Parameter
Alarming condition
Paw
Low
limit
Internally set alarm limit: PEEP+4 cmH2O
FiO2
High
limit
FiO2 exceeds the alarm limit for at least 20s.
Internally set alarm limit: Set value + max (7% or set value x 10%).
Low
limit
FiO2 is lower than the alarm limit for at least 20s.
Internally set alarm limit: set value - max (7% or set value x 10%).
Absolute FiO2 low limit: 18% O2
Sustained Airway
Pressure
Internally set alarm limit: PEEP+15 cmH2O
The alarm limit is exceeded for 15s continuously.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...