B-12
B.11 Compressor Specifications
Compressor Specifications
Input voltage
220 to 240 V
110 to 120 V
Input frequency
50/60Hz
60Hz
Input current
3A
6A
Output pressure range
300 to 450 kPa
Noise
Less than 50dB (A)
Continuous flow
≥
30 L/min at output pressure of 300kPa
Peak flow
Greater than 180 L/min for more than 0.8 seconds at one
barometric pressure
Dew point
Lower than the room temperature by 5
℃
at flow of 30 L/min
B.12 O
2
Sensor Specifications
O2 sensor
Output
9 to13 mV at 210 hPa O2
Range
0 to 1500hPa O2
100% O2 signal deviation
100±1%
Resolution
1 hPa O2
Expected service life
1.5 x 106 % for measurement (20°C)
0.8 x 106 % for measurement (40°C)
Response time ((21% air to 100% O2)
<15s
Linearity
Linear 0-100% O2
Operating temperature range
-20°C to +50°C
Temperature compensation
±2% of fluctuation at 0-40°C
Pressure range
50 to 200 kPa
Relative humidity
0 to 99%
100% O2 concentration output drift
Over one year of typical value <5%
Material White
ABS
Packaging Sealed
package
Period of validity
Not more than 13 months after unpacked (in compliance
with the conditions specified by the manufacturer)
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...