11-5
11.5 O2 Concentration Calibration
NOTE
Do not perform O2 concentration calibration while the unit is connected to a
patient.
Make sure that the system is Standby. If not, push the Standby key to enter
standby screen.
Calibrate the O2 sensor when the measured O2 concentration has a great deviation or when
the O2 sensor is replaced.
Follow these steps to calibrate the O2 sensor:
1. Make sure that Air and O2 supplies are connected.
2. Push the Menu key. Select [
Calibrate
] and select [
O2%
]. Select [
Calibrate
] on the
right side to start O2 sensor calibration. The message [
Calibrating
] is displayed.
3. During the calibration, if you select [
Stop
], the ongoing calibration will stop and the
message [
Calibration Stopped! Calibration is unfinished.
] is displayed.
4. After a successful calibration, the screen shows [
Calibration Completed!
]. Otherwise,
the message [
Calibration Failure! Please try again.
] is displayed. In this case, you
need to do the calibration again.
NOTE
If the calibration fails, check for technical alarm and troubleshoot it if there is.
Then do the calibration again. In case of repeated calibration failures, replace the
O2 sensor and do the calibration again. If it still fails, contact your service
personnel or us.
Handle and dispose of the O2 sensor according to your biohazard policies. Do not
incinerate.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...