B-11
B.10 CO
2
Module Specifications
CO
2
module
Measurement range and
accuracy
Measurement range
Accuracy
0 to 40 mmHg
±2 mmHg
41 to 76 mmHg
±5% of reading
77 to 99 mmHg
±10% of reading
Resolution 1
mmHg
Rise time
<330ms@100mL/min
<400ms@70mL/min
Delay time
<3s@100mL/min
<3.5s@70mL/min
Measured by using neonatal water trap and 2.5-meter long
neonatal sampling line.
<5s@100mL/min
<6.5s@70mL/min
Measured by using adult water trap and 2.5-meter long adult
sampling line.
Total system response time
<3.5s@100mL/min
<4s@70mL/min
Measured by using neonatal water trap and 2.5-meter long
neonatal sampling line.
<5.5s@100mL/min
<7s@70mL/min
Measured by using adult water trap and 2.5-meter long adult
sampling line.
Pump rate
70 mL/min and 100 mL/min optional.
The flow control accuracy is
±
15
﹪
of the set value or
±
15mL/min, whichever is greater.
Sidestream CO
2
alarm limits Range
Step
EtCO
2
high limit
(low limit + 2) to 99 mmHg
1 mmHg
EtCO
2
low limit
0 to (high limit - 2) mmHg
FiCO
2
high limit
1 to 99 mmHg
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...