7-2
NOTE
As required by the relevant rules and regulations, carbon dioxide concentration
should be monitored when the equipment is used on the patient. If your ventilator
is not configured with such monitoring function, use a monitor which complies
with the relevant international rules and regulations for carbon dioxide
concentration monitoring.
7.2 Prepare to Measure CO2
1. Attach the water trap to the water trap fixer and then connect the CO2 components as
shown below.
2. By default, the CO2 module is in measure mode. When the CO2 module is connected,
the [
CO2 Startup
] message is displayed on the screen.
3. After start-up is finished, the [
CO2 Warmup
] message is displayed. The CO2 module is
in ISO accuracy mode. If you perform CO
2
measurements during warm-up, the
measurement accuracy may be compromised.
4. After warm-up is finished, the CO2 module enters full accuracy mode.
It needs about 2 minutes after powering on the ventilator till the point that the CO2 module
enters full accuracy mode.
Water trap
Water trap fixer
Sampling line
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...