D-5
Check for leakage. Check if the expiratory trigger
sensitivity is too low.
Ctrl Module
Comm Error
L
Restart the ventilator. If the problem persists, contact your
service personnel.
Ctrl Module
Comm Stop
H
Ctrl Module
Selftest Error
H
Protection Module
Comm Error
L
Protection Module
Comm Stop
H
Protection Module
Selftest Error
H
Heating Module
Failure
L
Fan Failure
M
Pressure Sensor
Failure
H
Exp. Flow Sensor
Failure
H
Internal
temperature too
high
M
O2 Sensor Failure
M Replace the O2 sensor.
O2 Sensor
Unconnected
L
The O2 sensor is not connected. Re-connect the O2 sensor.
Please calibrate O2
sensor.
L
Calibrate the O2 sensor.
Please perform
flow calibration.
H
Calibrate the flow sensor.
Please perform
pressure
calibration.
H
Calibrate pressure.
Air Insp. Limb
Failure
H
The ventilator air limb supply is abnormal.
Check the gas supply and perform system test. If the alarm
is not cancelled, contact your service personnel.
O2 Insp. Limb
Failure
H
The ventilator O2 limb supply is abnormal.
Check the gas supply and perform system test. If the alarm
is not cancelled, contact your service personnel.
Insp. Gas Temp
Too High
H
The gas temperature exceeds 45
℃
. Restart the ventilator.
If the problem persists, contact your service personnel.
Buzzer Failure
L
The buzzer is faulty. Restart the ventilator. If the problem
persists, contact your service personnel.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...