11-10
NOTE
If the running time of the battery is too short after fully charged, the battery may
be damaged already or defective.
If obvious signs of damage are detected on the battery or the battery recharging is
failed, replace the battery and recycle it properly.
11.8.4 Battery Recycling
If obvious signs of damage are detected on the battery or the battery recharging is failed,
replace the battery and recycle it properly. Dispose of the battery in compliance with the local
laws regulating the disposal of such product.
WARNING
Do not disassemble batteries, dispose of them in fire, or short-circuit them. They
may ignite, explode, leak, causing personal injury.
11.9 Electrical Safety Inspection
NOTE
Perform electrical safety inspection after servicing or routine maintenance. Before
the electrical safety inspection, make sure all the covers, panels, and screws are
correctly installed.
The electrical safety inspection should be performed once a year.
11.9.1 Auxiliary Electrical Outlet Test
Verify the mains voltage is present at each auxiliary outlet when the ventilator is connected
with power.
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...