6-12
In CPAP (continuous positive airway pressure ventilation), the airway pressure is held at the
user-set positive pressure level throughout the ventilation cycle. The patient breathes
spontaneously and determines his own breathing frequency, tidal volume, and breath time.
The system starts apnea ventilation when detecting that the period of time when patient does
not perform continuous spontaneous breathing exceeds the preset apnea time.
Paw
Flow
Time
△
Papnea
Spontaneous breathing
Apnea ventilation
PEEP
Apnea vent cycle
I:E=1:2
Tslope
Apnea time
Time
In CPAP/PSV under non-NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Tslope
]:
Time
of
pressure
rising
3. [
△
Psupp
]:
Pressure support level
4 [
PEEP
]:
Positive
end-expiratory
pressure
5.
[
Exp
%
]:
Expiration
trigger
level
6. [
Δ
Papnea
]:
Pressure
of
apnea
ventilation
7. [
fapnea
]:
Frequency
of
apnea
ventilation
8. [
F-Trig
] or [
P-Trig
]: Inspiration
trigger
level
9. [
ATRC
]: Automatic tube resistance compensation
In CPAP/PSV under NIV, you need to set the following ventilation parameters:
1. [
O2%
]:
Oxygen
concentration
2. [
Tslope
]:
Time
of
pressure
rising
3. [
△
Psupp
]:
Pressure support level
4 [
PEEP
]:
Positive
end-expiratory
pressure
Summary of Contents for SynoVent E5
Page 1: ...SynoVent E5 Ventilator Operator s Manual...
Page 2: ......
Page 24: ...2 2 2 2 Equipment Appearance 2 2 1 Front View 4 1 2 3 5 6 7 8 9 10 11 12...
Page 46: ...3 18 FOR YOUR NOTES...
Page 60: ...4 14 FOR YOUR NOTES...
Page 68: ...5 8 FOR YOUR NOTES...
Page 92: ...6 24 FOR YOUR NOTES...
Page 118: ...9 12 FOR YOUR NOTES...
Page 154: ...A 8 FOR YOUR NOTES...
Page 169: ...B 15 Temperature C Life 1192 exp 2 0 0239 Temperature Life of 20 C Figure...
Page 170: ...B 16 FOR YOUR NOTES...
Page 184: ...D 8 FOR YOUR NOTES...
Page 195: ......
Page 196: ...P N 046 000991 00 10 0...