80
47LLC OPERATING MANUAL EM-9505-B Rev.4
8. OPERATION
Make sure that 0.0 – 300.0 kg is correctly indicated with the actual load applied.
IMPORTANT
Before operating, make sure that the wiring is correct, the input, the excitation voltage and the power supply are within
the specification range.
1
With no load applied or the empty vessel put on the strain
gauge, make sure that 0.0 kg is indicated.
NOTE
■
WHEN THE INDICATION Is sHIFTED…
•
When the indication is shifted without any load, not using the vessel, carry out
the Teach Calibration. (Refer to 6. SETTING SCALING VALUES WITH ACTUAL
LOAD.)
•
When the indication is shifted with the empty vessel put on, adjust the tare. (Refer
to 7. ADJUSTING TARE.)
Blinking
Blinking
■
WHEN THE FOLLOWING Is INDICATED…
•
When ‘S.ERR’ is indicated, the input is not applied correctly. Check the input wiring, equipment, signal and excita-
tion voltage. When ‘Min’ indicator blinks, the input signal is under the specification voltage. And when ‘Max’ indicator
blinks, the input is over the specification voltage.
■
ALARM INDICATORs
•
The status of the alarm indicators depends on the alarm setpoints. The above display example shows ‘P’ indicator on.
2
With 300 kg load applied, make sure that 300.0 kg is indicated.
*
1 Display depends on the settings and input.