64
47LLC OPERATING MANUAL EM-9505-B Rev.4
6.3.1 OPERATING PROCEDURE
Procedures to set the load ratio are described here, not with the rated load 300 kg but with a weight 30 kg, 10% of the load.
NOTE
The left figure shows a display example. The display
depends on the settings.
1
Confirm the wiring, and turn on the power.
•
All the indications turn on for approximately 1 second and then
the display moves on to Measuring Mode.
NOTE
Indication ‘S.ERR’ may blink, which shows the input out of the measur-
ing range and does not show the unit failure.
2
Hold down Scale/
↑
button for 3 seconds or more to move on to
Scaling Setting Mode.
•
The input type is indicated.
•
‘Zro’ and ‘Spn’ indicators turn on.
NOTE
Some alarm indicators turn on with the function indicators. The alarm indication is due to the last status before mode
transition held but does not show the unit failure.
3,6
2,3,
6
4,5
5
■
Immediately after power on (all indicators on)
■
Measuring Mode
*
1 Display depends on the settings and input.