![background image](http://html1.mh-extra.com/html/m-system/47llc/47llc_operating-manual_3126851077.webp)
77
47LLC OPERATING MANUAL EM-9505-B Rev.4
7.2 TARE ADJUsTMENT VIA BUTTON CONTROL
Procedures to adjust the vessel weight 30.0 kg via front button control are described here with the input scaling 0 – 10 mV
DC and display scaling 0.0 – 300.0 kg.
7.2.1 OPERATING PROCEDURE
NOTE
The left figure shows a display example. The display
depends on the tare adjustment.
1
Hold down Scale/
↑
button for 3 seconds or more to move on to
Scaling Setting Mode.
•
The input type is indicated.
•
‘Zro’ and ‘Spn’ indicators turn on.
2
Press Alarm/
↓
or Scale/
↑
button to go to the tare adjustment.
•
The tare adjustment is indicated.
•
‘Zro’, ‘Tch’ and ‘Fnc’ indicators turn on.
3
Press Up button to go to the teach calibration (TARE) setting.
•
‘Tch’ indicator starts blinking.
IMPORTANT
The value added to the input scaling value A is indicated. In this example, the vessel weight is equivalent to 10% of the
display scaling. The strain gauge output voltage is 1 mV, equivalent to 10% of 10 mV. Therefore ’01.00’, 1 mV added to
the input scaling value A, is indicated.
2,5
1,2,
5
3,4
Blinking