111
47LLC OPERATING MANUAL EM-9505-B Rev.4
4
Press Alarm/
↓
or Scale/
↑
button to apply the new setting.
•
And the next parameter setting is indicated.
NOTE
•
Press Alarm/
↓
button, and the low-end cutout value will be indicated within the range of ‘Z000’ to ‘Z999’ depending on
the setting. When low-end cutout OFF is selected, the brightness ‘C 1’, ‘C 2’, ‘C 3’, ‘C 4’ or ‘C 5’ will be indicated depend-
ing on the setting.
•
Press Scale/
↑
button, and the moving average sampling No. ‘AOFF’, ‘A 2’, ‘A 4’, ‘A 8’, ‘A 16’, ‘A 32’ or ‘A 64’ will be indicated
depending on the setting.
5
■
TO GO ON TO SET THE LOW-END CUTOUT VALUE,
Skip to Step 2 in “13.2 LOW-END CUTOUT VALUE”.
■
TO qUIT,
Hold down Alarm/
↓
or Scale/
↑
button for 1 second or more to return to Measuring Mode.