59
47LLC OPERATING MANUAL EM-9505-B Rev.4
4
Press Up button to go to the teach calibration (ZERO) setting.
•
The present input is indicated.
•
‘Tch’ indicator starts blinking.
5
Without any load (input 0%), press Up button to register the
present value.
•
The teach calibration (ZERO) is registered.
•
‘Tch’ indicator turns ON.
IMPORTANT
•
Confirm that the input signal is stable before pressing Up button.
•
Set within the setting range.
NOTE
•
‘00.00’ is a display example. The registered input value is indicated.
•
‘Min’ and ‘Max’ indicators start blinking when the set value is within invalid range. Return the setting within the valid
range.
6
■
TO GO ON TO sET THE DIsPLAY sCALING VALUE
A,
Skip to Step 3 in “6.2 STEP 2. DISPLAY SCALING VALUE A (ACTUAL LOAD)”.
■
TO qUIT,
Hold down Alarm/
↓
or Scale/
↑
button for 1 second or more to return to Measuring Mode.
NOTE
■
IF THE FRONT BUTTONs ARE LEFT UNTOUCHED…
•
The display goes back automatically to Measuring Mode after the specified time period (default: 15 sec.) in the steps
other than Step 4.
•
This time period (automatic return time) is configurable. (Refer to 15. GOING BACK AUTOMATICALLY TO MEASUR-
ING MODE.)
■
TO ABORT A sETTING…
•
Hold down Max/Min button for 1 second or more in Step 4 to return to Measuring Mode without applying the last
changes.
•
If you get lost in a setting mode, you can execute initialization. (Refer to 21.2 INITIALIZING SETTING VALUES.)
Blinking