![background image](http://html1.mh-extra.com/html/m-system/47llc/47llc_operating-manual_3126851149.webp)
149
47LLC OPERATING MANUAL EM-9505-B Rev.4
22.4.2 sCALING sETTING MODE
NOTE
•
The display depends on the specifications, settings and input.
•
The analog output parameters are not indicated with no-DC-output type.
•
The input type is for indication only, not for setting.
Measuring Mode
for ≥ 3 sec.
for ≥ 1 sec.
or
Input type
Analog output
function mode
DISP, SCAL
No decimal point,
10
-1
to 10
-4
Decimal point position
Measurement
stopped
Measurement
started
Scaling Setting Mode
(depending on input strain gauge codes)
Analog output 0% adj.
UP (increasing),
DOWN (decreasing)
Analog output 100% adj.
UP (increasing),
DOWN (decreasing)
Teach calibration (ZERO)
Display scaling value A
-19999 to 19999
Input scaling value A
-30.00 to 30.00
(depending on input strain gauge codes)
-27.00 to 60.00
(depending on input strain gauge codes)
Input scaling value B Teach calibration (SPAN)
Display scaling value B
-19999 to 19999
Teach calibration (TARE)
Excitation voltage
2.5 V, 5 V, 7.5 V, 10 V
Load ratio
10.0 to 100.0%
-24.00 to 24.00
(depending on input strain gauge codes)
Tare adjustment