58
47LLC OPERATING MANUAL EM-9505-B Rev.4
6.1.2 OPERATING PROCEDURE
NOTE
The left figure shows a display example (default value
of input strain gauge code S2). The display depends on
the settings. Refer to 6.1.1 LIST FOR INPUT SCALING
VALUE A (ACTUAL LOAD) for details.
1
Confirm the wiring, and turn on the power.
•
All the indications turn on for approximately 1 second and then
the display moves on to Measuring Mode.
NOTE
•
Warm up the unit for 10 minutes or more.
•
Indication ‘S.ERR’ may blink, which shows the input out of the meas-
uring range and does not show the unit failure.
2
Hold down Scale/
↑
button for 3 seconds or more to move on to
Scaling Setting Mode.
•
The input type is indicated.
•
‘Zro’ and ‘Spn’ indicators turn on.
NOTE
Some alarm indicators turn on with the function indicators. The alarm indication is due to the last status before mode
transition held but does not show the unit failure.
3
Press Alarm/
↓
or Scale/
↑
button to go to the input scaling
value A setting.
•
The input scaling value A is indicated.
•
‘Zro’ and ‘Tch’ indicators turn on.
3
2,3
4,5
■
Immediately after power on (all indicators on)
■
Measuring Mode
*
1 Display depends on the settings and input.