Troubleshooting
5
Spring damps suddenly after
unlocking.
Spring is off level, check bubble
levels for proper level.
Spring is out of range and is
hitting upper or lower physical
limit of the spring.
Horizontal flexure ( delta rod )has
been damaged and needs
replacement.
The lock on the controller is
locked.
Gravity is extremely noisy (mGal)
It is possible that there is an
earthquake event occurring
somewhere on earth. The
Superspring is very sensitive to
earthquakes no matter where they
occur. The gravity will continue to
be noisy for hours or even days
when the earthquake event is close
to the gravimeter location.
Table 5-3
Superspring Problems and Solutions
5-9