Adjustment and Maintenance
4
(electrical feedthrough connector), then remove the electrical feedthrough from
the service ring.
!
MAKE SURE TO NOTE THE COLOR CODING ON THE
CONNECTORS FOR LATER RECONNECTION.
Remove the travel lock knob by removing the flat-head screw from the center of
the knob and sliding the knob off the shaft. The knob is comprised of three
pieces: the brass knob, the control spring retainer, and the internal spring. With
the knob removed, the two screws holding the travel lock knob assembly are
exposed. Remove these and slide the assembly out of the service ring (the travel
lock to service ring interface is sealed with Teflon sealer).
The service ring and bottom flange can be taken off together by removing the
three screws securing the base plate of the Superspring structure to the bottom
flange.
4.3.1.3.Replacing the Coil (Linear Actuator)
The permanent magnet assembly of the coil is attached to the base plate, and the
voice coil
pusher
is attached to the emitter/detector block of the support
structure.
Remove the service ring as described previously.
Remove the base plate by loosening the three 3mm barrel clamp screws that
secure the rods to the plate. Carefully lift rod assembly off bottom plate.
Remove two shoulder screws that hold travel lock fork on (from underneath)
and slide travel lock fork assembly off main rod. Unhook three support springs
from O-rings and remove lower triangular spring plate. This plate is held on by
four 3-mm screws. Be careful not to cut or score the O-rings. Note orientation of
emitter-detector block and remove block (six 3mm screws in deep counter bore
holes). Remove the permanent magnet assembly from the base plate.
4-14
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......