Troubleshooting
5
Hum at the top of the drop.
Hold servo gain needs to be
reduced.
Shaft encoder is dirty and needs to
be blown clean with high pressure
air.
Upward or downward trend in gravity
values (time series plot).
Ion pump is not turned on.
Time or date on PC controller
have been set incorrectly.
Tide correction flag has not been
set in the Process | Setup |
Control.
Latitude or longitude have been
entered incorrectly in Process |
Setup | Information.
Set up on non-rigid floor and
verticality of the beam is drifting.
Single clicking sound during lift and
drop.
Drive belt developing fracture and
needs to be replaced.
Stops after one drop and resets itself
Drive belt could be loose and may
need to be tightened.
Multiple clicking sounds during lift
and drop.
Drive bearing needs to be
replaced.
Cart drives to the top very fast
No shaft encoder signal. Check
cable connection. Check that the
cart voltage changes as you
manually lift the mass (using a hex
wrench).
Very bad sound during drop
Sphere signal is not present
because of the cable is not
connected or a there is a bad
cable/connector. Look for a sphere
signal that changes as the mass is
travel locked.
5-3
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......