Troubleshooting
5
Sharp spikes (faster than 1ms) on the
sphere signal during drop with a bad
sound during drop
Bad cable/connector
Many bad drops occur with extremely
high residuals. [High number of drop-
outs]
The cart travel is not aligned with
the vertical. The bubble levels on
the tripod need to be reset or the
dropping chamber is not sitting in
the tripod tray correctly.
Dropper guide rods are not
straight. Can be due to snubbers
tightened asymmetrically or
because of stress from bad
alignment of rods in the top ring.
Table 5-1
Dropping Chamber Problems and Solutions
5-4