Adjustment and Maintenance
4
Figure 4-2 APD alignment adjustment access
23. Adjust the X-Y stage of the APD to maximize the voltage at the “analog” BNC
on the interferometer base. This voltage is negative. Adjust the stage to
obtain smallest absolute value. A digital voltmeter or oscilloscope is
recommended for this procedure. In both the X and Y directions, there
should be a “plateau” of stable, less negative voltage. Adjust both the “X”-
screw (accessed through a hole on the rear of the I.B.) and “Y”-screw
(accessed by “grabbing” the knurled knob of the XY stage) so that the APD is
in the center of both “plateaus”.
24. Replace the rear plate on the interferometer base and tighten all screws.