How to Set Up and Run the FG5
3
41. Connect the gray ribbon cable from the “rear of the patch panel to the
connector on the computer. Note the location of Pin 1 (white arrow) in both
cases.
42. Connect the BNC cable from the interferometer base “TTL” connector to the
“FRINGES” connector on the computer time interval card, identified as
“Channel A” in the Time Interval Analyzer (TIA) operating manual.
43. Connect the BNC cable from the “10 MHz” connector on the power supply to
the “CLK” connector on the computer time interval card, identified as “EXT
CLK” in the TIA operating manual.
44. Connect the BNC cable from the TRIG OUT connector on the dropper
controller to the “TRIG” connector on the computer time interval card,
identified as “EXT ARM” in the TIA operating manual.
45. Connect the barometer cable from the power supply to Channel 4 on the
patch panel.g
46. For the WEO Model 100 laser:
Connect the BNC cable from the OUTPUT connector on the front panel of the
laser controller to Channel 3 of the patch panel . Make sure the meter select
switch on the laser controller is set to the “1F” position.
47. For the Model ML-1 Laser: Connect the LASER LOCK on the patch panel to
the REMOTE BNC on the back of the ML-1 controller (this allows the laser
mode to be switched from “red” to “blue”). Connect the DIG C5 output on
the patch panel to REMOTE2 BNC connector on the back of the ML-1
controller (this enables the warmup/lock mode) .
48. Connect the BNC cable from the SPHERE OUT connector on the superspring
controller to the Channel 1 of the patch panel.
49. Connect the BNC cable from the METER MONITOR connector on the ion
pump power supply to Channel 2 of the patch panel. Set the meter select
switch to the 10
-4
µ
A scale.
50. Connect the BNC cable from the TRIG OUT connector on the patch panel to
the TRIG EXT IN connector on the dropping chamber controller.
3-10
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......