Checklists and Logs
Appendix D
9.2. FG5 Setup
FG5 Setup Checklist
Locate
the
reference
mark.
Place the electronics case in a convenient location.
Check AC line voltage. Connect main AC power cable.
Turn on main AC and DC power switches.
Turn on laser power.
Place superspring tripod over reference mark.
Rough level superspring tripod.
Measure and record lower reference height.
Place superspring on tripod and clamp in position.
Level superspring using precision level vials.
Remove superspring dust cover and place interferometer on
superspring. Tighten four knobs to lock interferometer to
superspring.
Remove dust cover from interferometer base.