Checklists and Logs
Appendix D
Make sure new bearings are cleaned and lubricated with Krytox
LVP low vacuum grease.
Reassemble in reverse order (make sure bowed snap ring is
oriented correctly and fully seated in groove).
Table 9-11 Replacing Shaft Bearings (Drive Pulley)
9.4.9.Replacing Shaft Bearings (Top Pulley)
Replacing Shaft Bearings (Top Pulley)
Remove dropping chamber cover.
Remove drive belt.
Remove snap rings on upper bump stop and slide down shaft.
Remove retaining rings on top of upper pulley yoke.
Remove pulley (slide down).
Remove bowed retaining ring from short end of pulley shaft (note
orientation).
Remove retaining ring from other end of shaft.
Remove pulley from bearing mounting yoke.
Push bearings out of yoke.
Make sure new bearings are cleaned and lubricated with Krytox
LVP low vacuum grease.
9-14
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......