Checklists and Logs
Appendix D
Place superspring alignment fixture on top superspring window.
Verify beam alignment through superspring and adjust, if
necessary.
Adjust mirror # 2 to align test and reference beams in telescope.
Adjust twiddler to align test and reference beams in fringe viewer.
Remove dropping chamber and tripod from interferometer base.
Block test beam and remove top plate of interferometer base.
Adjust APD X-Y stage to maximize voltage at “analog” BNC
connector on interferometer base (largest negative voltage).
Replace top lid of interferometer base and tighten all screws.
Table 9-14 Interferometer Alignment
9-17