Troubleshooting
5
5.2.6.ENVIRONMENTAL SENSORS
PACKAGE
The environmental sensors package is an optional system for the FG5. Therefore,
if there is a major failure, it is possible to continue observations. Occasionally,
the operator can obtain temperature and pressure data from an alternate source.
If the pressure data is accurate enough (
±
1 mb or better), local barometric
pressure attraction corrections accurate to better than one
µ
Gal can still be
computed and applied during post processing. Possible problems and solutions
pertaining to the environmental sensors package are shown in Table 5-6.
Problem
Solution
Gravity is 300
µ
Gal too high
The system is not reading the
barometer. Make sure the
barometer is hooked up correctly.
Also make sure that the voltage
range for the analog input (Process
| Setup | System | A2D Card) is
set to Unipolar, 10 V, with an
offset of 537.5 and a multiplier of
1.25.
Table 5-6
Environmental Sensors Package Problems and Solutions
5-12
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......