Adjustment and Maintenance
4
4.1.1.4.Adjusting the Drive Belt Tension
Adjust belt tension using the set screw on the top pulley assembly, located on the
top rod ring. Tighten the belt adjustment screw until the slack has been taken
out and the belt is straight. Then tension the belt by tightening the screw
approximately three turns. If you are uncertain of the proper tension, the screw
can be tightened until the tension spring is just short of coil bind. The belt can
also be tensioned using the torque required to slip the pulley on the belt to
determine belt tension. Use a torque wrench to manually drive the motor shaft
(drive assembly). Set the torque wrench in the 6-7 inch-lb. range and tighten the
set screw until the belt slips on the pulley when rotated into the lower stop.
4.1.1.5.Replacing the Ferrofluidic Vacuum Feedthrough
The chamber must be vented and opened for this procedure. If possible, use dry
nitrogen to vent the chamber.
Loosen the English 4-40 clamp screws on the Helical coupling between the motor
and the Ferrofluidic vacuum feedthrough by reaching through the access hole in
the motor mount. Remove the three ¼-28 English screws which attach the motor
mount to the Conflat vacuum flange on the service ring. Remove the motor
mount assembly (including motor, Helical coupling, travel lock plate, and
encoder) from the Conflat vacuum flange, leaving the Conflat flange and
Ferrofluidic vacuum feedthrough attached to the service ring.
Inside the service ring, loosen the socket head clamp screw on the Helical shaft
coupler where it attaches to the lower drive pulley shaft. Remove the three
remaining Conflat mounting screws. Remove the Conflat with Ferrofluidic
feedthrough attached.
Remove the Helical coupling from the feedthrough, and unscrew the
feedthrough from the Conflat flange.
Lubricate the O-ring on the new Ferrofluidic vacuum feedthrough with a light
coat of Apiezon L grease. Use pliers with padded jaws (e.g. blue pad) to gently
tighten the feedthrough to the Conflat flange.
4-5
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......