Checklists and Logs
Appendix D
Remove emitter-detector block (note orientation).
Remove permanent magnet assembly from base plate.
Remove coil from emitter-detector block.
Reverse procedure for reassembly.
Table 9-18 Replacing Superspring Coil
9.6.5.Replacing Superspring Flexures
Replacing Superspring Flexures Checklist
Remove Superspring cover.
Put dust cover over top main tube.
Remove service ring.
Place hard foam or rubber shim between top rod ring and top
triangular plate.
Fasten plates together with cable tie.
Loosen clamps holding rods to bottom plate.
Disconnect wires from pre amp circuit board.
Lift rod assembly off bottom plate.
9-20
Summary of Contents for Micro-g LaCoste FG5
Page 7: ...Table of Figures This Page is Intentionally Blank vii ...
Page 8: ......
Page 13: ...Design Components and Function 2 Figure 2 3 Side view of the dropping chamber 2 3 ...
Page 23: ...Design Components and Function 2 Figure 2 9 The Superspring 2 13 ...
Page 31: ...Design Components and Function 2 Figure 2 12 Rotation Monitor 2 21 ...
Page 32: ...Design Components and Function 2 This Page is Intentionally Blank 2 22 ...
Page 36: ...How to Set Up and Run the FG5 3 Figure 3 1 Superspring Interferometer Setup 3 4 ...
Page 56: ......
Page 59: ...Adjustment and Maintenance 4 4 3 ...
Page 78: ...Adjustment and Maintenance 4 4 22 ...
Page 106: ......
Page 138: ...Checklists and Logs Appendix D Measure gap between two levers near ZPM 9 24 ...
Page 143: ...Checklists and Logs Appendix D This Page is Intentionally Blank 9 29 ...
Page 144: ......