Checklists and Logs
Appendix D
9.6.8.Adjusting Superspring Micro-
Switches
Adjusting Superspring Micro-Switches Checklist
Adjust trip positions by changing position of set screws in actuator
arms.
Adjust so limit switches shut off ZPM when gap between focus
lever and fulcrum plate is 4-8 mm.
Table 9-22 Adjusting Superspring Micro-Switches
9.7. Packing the FG5
Packing the FG5 Checklist
Unplug cables from electronics rack and components.
If leaving the chamber under ion pump control, leave ion pump
cables attached from power supply to dropping chamber.
Put remaining cables in zippered pouch (inside of rear electronics
case lid).
Secure lids to electronics case.
Unplug power cords and printer cable from computer.
9-26