Operation
5
3. The blade holder base
can be moved back and forth on the microtome base
plate. This allows bringing the two-in-one blade holder E to optimal sectioning position in
relation to the specimen. To secure the blade holder base, turn the clamping lever
clockwise.
5.2.2 Inserting the two-in-one blade holder E
Fig. 54
7
8
9
10
1. Loosen the eccentric bolt
using an Allen key No. 4
.
2. Place the two-in-one blade holder E
with the underside groove onto the T-piece
3. To clamp, retighten the eccentric bolt
.
5.3
Adjusting the clearance angle
Fig. 55
1
2
3
4
5
6
The index marks (0°, 5° and 10°) for adjustment of the clearance angle
are located on the
right side of the two-in-one blade holder E
the right side of the blade holder base
which serves as a reference point when adjusting
the clearance angle.
until the two-in-one