30
Version 1.1, Revision B
Setting up the instrument
4
4.
Setting up the instrument
4.1
Installation site requirements
• Stable, vibration-free laboratory table with horizontal, flat table top, as far as possible vibration-free
ground.
• No other instruments nearby which might cause vibrations.
• Room temperature consistently b 18 °C and + 30 °C.
• Obstruction-free access to the handwheel.
• To ensure proper function of the instrument, it must be set up while maintaining a minimum distance
of 10 cm from walls and furniture.
• The instrument must be installed in a place that ensures an easy disconnection from the power
supply. The power cable must be in a place that can be easily reached.
Danger
Explosive hazard.
Death or Serious injury of persons, and/or property damage.
• Never operate the instrument in rooms with an explosive hazard.
4.2
Standard delivery - packing list
The HistoCore AUTOCUT is available in two configurations.
HistoCore AUTOCUT configuration: 14 9AUTO00C1
Quantity
Part description
Order No.
1
HistoCore AUTOCUT basic instrument
14 0519 56472
1
Anti-static waste tray
14 0517 56237
1
Fine-directional fixture for specimen clamp
14 0502 37717
1
Quick clamping system
14 0502 37718
1
Universal cassette clamp
14 0502 37999
1
Blade holder base
14 0502 55546
1
2-in-1 Blade holder E
14 0502 54497
HistoCore AUTOCUT configuration: 14 9BIO000C2
Quantity
Part description
Order No.
1
HistoCore AUTOCUT basic instrument
14 0519 56472
1
Standard waste tray
14 0518 56458
1
Fine-directional fixture for specimen clamp
14 0502 37717
1
Quick clamping system
14 0502 37718
1
Universal cassette clamp
14 0502 37999
1
Blade holder base
14 0502 55546
1
2-in-1 Blade holder E
14 0502 54497
HistoCore AUTOCUT basic instrument includes the following delivery list.