73
HistoCore AUTOCUT
Optional Accessories
6
Fig. 75
1
2
3
4
5
6
Clamping of flat, angular specimens
To clamp angular specimens, replace the long
setscrew
• Unscrew the long setscrew
to
the left with an Allen key No. 4
• Screw the short setscrew
in the
hole.
• Place the sample
movable jaw
and the fixed jaw
.
• To clamp the sample, screw the movable jaw
.
• Insert the foil clamp in the standard specimen
clamp as shown.
clockwise to clamp the foil clamp in the
standard specimen clamp.
6.2.4 Universal cassette clamp
Warning
Insufficient frame conditions for sectioning.
Specimen damage or poor sectioning results, for example, sections alternating in thickness,
compressed, folded or stripped sections.
• Do not continue sectioning if you recognize insufficient sections results.
• Make sure that all prerequisites for appropriate sectioning are met. Refer to the troubleshooting
section of these Instructions for Use for further details.
• If you do not have sufficient knowledge in regards of troubleshooting insufficient sectioning
results consult persons who do have that knowledge, for example, application experts of Leica
Biosystems.