81
HistoCore AUTOCUT
Optional Accessories
6
Warning
The knife/blade is installed before the knife/blade holder and the knife/blade holder base are installed
on the instrument.
Serious injury of persons due to contact with extremely sharp knives and/or blades.
• Prior to inserting the knife/blade, both the knife/blade holder and the knife/blade holder base must
be installed on the microtome.
Inserting the knife
on the right and left of the knife holder forward in opposite
directions, lowering the knife support bar to the lowest possible position, thus ensuring that the knife
edge will not be damaged when inserting the knife.
• Unscrew the clamping screws
(rotate counterclockwise) until the knife can safely be
inserted.
• Hold the knife
at the knife base and carefully insert it in the holder from the side as
shown with the cutting edge facing upward.
Fig. 88
1
2
Either side of the profile c knife
can be inserted, whereas the shining facet of profile d
should face the operator. Putting the knife wrongly will damage the specimen and
the knife.
Knife height adjustment
When adjusting the knife height, the knife edge should be positioned as exactly as possible in the actual
center of rotation of the knife holder. The lay-on edge
of the rear clamping chucks serves
as a reference position for correct knife height adjustment.
uniformly and backwards until the knife blade is parallel to the
evenly screw the two knife clamping screws
(rotate clockwise).
Lateral repositioning of the knife
• Loosen the clamping screws
by turning them counterclockwise.
• Push the knife
to the left or right as required.
• After the knife has been repositioned, adjust the knife height
P. 81 – Knife height adjustment)
then tighten the clamping screw
which is located on the side to by turning it clockwise
.