28
Version 1.1, Revision B
Instrument Components and Specifications
3
3.3
Technical Data
General
Nominal supply voltages
100/120/230/240 V AC ±10%
Nominal frequency
50/60 Hz
Maximum power consumption
100 VA
Protection class
(according to IEC-1010, UL 3101, EN 61010)
I
Power fuses
2 x T 3.15 A
Pollution degree
(according to IEC-1010, UL 3101, EN 61010)
2
Overvoltage category
(according to IEC-1010, UL 3101, EN 61010)
II
IP protection class
IP20
Operating temperature range
+18 °C to +30 °C
Operating relative humidity
20 % to max. 80 % non-condensing
Operating altitude
Up to 2000 m above sea level
Transport temperature range
-29 °C to +50 °C
Storage temperature range
+5 °C to +50 °C
Transport and storage relative humidity
10 % to max. 85 % non-condensing
Dimensions and weights
Basic instrument
Width (including the handwheel and the
coarse feed wheel)
477 mm
Depth (including the section waste tray)
620 mm
Height (without the top tray)
295 mm
Weight (without accessories)
Approx. 40 kg
Volume of the section waste tray
1400 ml
Control panel
Width
119 mm
Depth
193 mm
Height
50 mm
Height (in inclined position)
81 mm
Weight (net)
approx. 0.6 kg
Microtome
Sectioning thickness setting
Section thickness setting range
0.50 - 100 µm