23
HistoCore AUTOCUT
Safety
2
Warning
Misinterpretation of the yellow LED in the
M-STOP
field.
Serious injury and/or property damage.
• Operate the instrument only after you have completely understood the meaning of the 3 possible
LED indications. Extinguished light: the instrument is ready to use (no brake engaged); blinking
light: the electronic brake is engaged, but mechanical locking system is deactivated; permanent
light: the mechanical locking system is activated.
• The handwheel is securely locked only when the handwheel locking mechanism
the 12 o'clock position.
Centering the handle
Fig. 4
1
1
2
2
For safety reasons, the handle of the handwheel
must always be centered while in motorized
sectioning mode.
• Activate the handwheel locking mechanism.
• To center the handwheel, lightly pull the handle
outwards and swivel it into the
center of the handwheel
• Release the handle and it will then be fixed in
the center of the handwheel.
2.3.3 Safety guard on the knife/blade holder
Each knife holder is equipped with a tightly mounted safety guard
. This makes it possible to cover completely the cutting edge in every knife or blade
position.
Fig. 5
1
2
2
1
2
2
Fig. 6
Knife holder N
N can be easily positioned via two handles
. To cover the knife edge, push both
cover strips of the safety guard to the center.
Knife holder NZ
NZ can be easily positioned via two handles
. To cover the knife edge, push both
cover strips of the safety guard to the center.