Version 1.1, Revision B
Operation
5
2. Move the two-in-one blade holder E until the index mark of the desired clearance angle
coincides with the reference line on the blade holder base. Enlarged detail
a clearance angle setting of 5°.
Tips
The recommended clearance angle setting for the two-in-one blade holder E is approx.
2.5° - 5°.
3. Hold down the two-in-one blade holder E in this position and retighten the screw
for clamping.
5.4
Inserting the universal cassette clamp
Fig. 56
1
1
2
3
4
5
6
7
There are two specimen holder fixtures with orientation and one without, please see
. The specimen orientation allows for simple position correction of the specimen
surface when the specimen is clamped into place. You can use the quick clamping system
to hold all available accessory specimen clamps (for more information, see
To do so, proceed as follows:
to the upper end position by turning the handwheel
and engage the handwheel locking mechanism.
2. To release the clamping system, turn the screw
counterclockwise using an Allen key No. 4
3. Push the guide
of the universal cassette clamp
from the left into
the quick clamping system
as far as it will go.
4. To clamp the cassette clamp turn the screw
clockwise as far as it will go.
Tip
Since all specimen clamps available as accessories are equipped with the same kind of guide on the
back, they are inserted in the same way described here using the example of the cassette clamp.