69
HistoCore AUTOCUT
Optional Accessories
6
6.1.2 Directional fixture for specimen clamps
Fig. 69
1
2
3
4
5
6
7
8
turning it counterclockwise.
• Completely unscrew the thrust piece
with a flat-tip screwdriver and pull it out
with spring
• Completely unscrew the setscrews
• Install the directional fixture for specimen clamps
by tightening the screws in the bores
(2 screws are accessible through the bores)
and screw them in using an Allen key No. 3. Insert
the screws
in the bores as shown
and screw them in using an Allen key No. 3.
• Insert the spring
with the flatter side into the thrust piece
. Completely screw in the thrust
piece with a flat-tip screwdriver.
• Completely screw in the setscrews
.
6.1.3 Fine-directional fixture for specimen clamps
Fig. 70
1
3
4
2
• Before the fine-directional fixture for specimen
clamps can be mounted, loosen 4 screws
(Allen key No. 3) and carefully
remove the fixture for specimen clamps from
and
the Allen key No. 3, fasten the baseplate to the
• Now, screw the fine-directional fixture for
specimen clamps with the 4 screws
and the Allen key No. 3 onto the object
head.