35
HistoCore AUTOCUT
Setting up the instrument
4
4.4
Assembling the handwheel
Tip
The handwheel has to be assembled before attempting to use the instrument. The necessary parts
and tools can be found in the tool kit supplied in the delivery.
Fig. 15
4
3
2
1
5
6
7
The feather key
is loosely placed in the handwheel shaft
and fixed in place
with a cable tie during transport.
Caution
The feather key is lost when mounting the handwheel.
The instrument cannot be used, and it may cause the delay of diagnosis.
• Before assembling the handwheel, check the feather key and make sure that it is in the
handwheel shaft.
2. Place the spacer
3. Tighten the screw located in the center hole of the handwheel with an Allen key No. 4
.
4. Remove the cover foil from the self-adhesive cover disk
and fix the cover disk on
the handwheel.
4.5
Electrical connections
Warning
Connecting the instrument to a non-grounded socket, or use an extension cord.
Risk of electrical shock which causes injury to persons, or delayed diagnosis.
• The instrument must be connected to a grounded socket.
• Do not use an extension cord.