12. Vector Display
12-9
12.6
Switching between the Vector, 5 Bar, and Phase Difference Displays
To switch between the vector, 5 bar, and phase difference displays, follow the procedure
below.
Procedure
VECTOR
→
F•6 DISPLAY
Settings
VECTOR:
The vector display is shown. This is the default setting.
5BAR:
The 5 bar display is shown.
EXTPHASE: The phase difference between an SDI signal and an external sync signal is
displayed.
12.6.1
5 Bar Display Explanation
In the 5 bar display, the positive and negative peak levels are displayed simultaneously.
The levels are typically displayed in cyan, but portions that exceed their limits are displayed
in red.
To show the 5 bar display, set F•6 DISPLAY to 5BAR.
Figure 12-11 5 bar display
Table 12-2 5 bar display explanation
No. Item
Explanation
1
Y
The luminance signal level is displayed here. Levels less than 0 % or greater
than 100 % are displayed in red.
2
G, B, and R
The GBR signal levels of the converted Y,C
B
,C
R
signal are displayed. The levels
that fall outside of the range you set using GAMUT ERROR in the status display
are displayed in red.
Reference: Section 14.6.4, “Setting Gamut Error Detection Levels”
3
CMP
The pseudo-composite signal level of the converted Y,C
B
,C
R
signal is displayed.
(The blanking interval is not included.) Levels that fall outside of the range you
set using COMPOSIT GAMUT in the status display are displayed in red.
Reference: Section 14.6.5, “Setting Composite Gamut Error Detection Levels”