14. Status Display
14-31
14.6.5
Setting Composite Gamut Error Detection Levels
To set composite gamut error detection levels, press F•2 COMPOSIT GAMUT in the status
menu.
The settings that you make here also apply to the CMP bar in the 5 bar display.
Reference: Section
12.6.1, “5 Bar Display Explanation”
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•2 COMPOSIT GAMUT
→
Figure 14-20 COMPOSIT GAMUT menu
• Setting the Upper Limit
To set the composite gamut error upper limit, follow the procedure below. An error is
detected when the level of the pseudo-composite signal that is converted from the SDI
input signal exceeds the specified value.
If you press F•D, the upper limit will be reset to its default value of 135.0.
Procedure
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•2 COMPOSIT GAMUT
→
F•1 C.GAMUT UPPER %
Settings
Selectable range: 90.0 to 135.0 (The default setting is 135.0.)
• Setting the Lower Limit
To set the composite gamut error lower limit, follow the procedure below. An error is
detected when the level of the pseudo-composite signal that is converted from the SDI
input signal goes below the specified value.
If you press F•D, the lower limit will be reset to its default value of –40.0.
Procedure
STATUS
→
F•5 ERROR CONFIG
→
F•4 ERROR LEVEL
→
F•2 COMPOSIT GAMUT
→
F•2 C.GAMUT LOWER %
Settings
Selectable range: –40.0 to –20.0 (The default setting is –40.0.)