14. Status Display
14-5
Table 14-2 Event log explanation
No. Item
Explanation
1
SAMPLE No.
The total number of events (a number from 0 to 1000) is
displayed.
2
Logging status
“NOW LOGGING” appears here when events are being
logged. “LOGGING STOPPED” appears when logging is
stopped.
You can start and stop event logging by pressing
F•2
LOG.
Reference: Section 14.2.3, “Starting Event Logging”
3
Event numbers
Events are assigned numbers in order of their occurrence.
The most recent event appears at the top of the list. To view
earlier events, turn
F•D
to the right.
You can display a maximum of 1000 events. To set whether
or not events after the 1000th event are written over earlier
events, press
F•4
LOG MODE.
Reference: Section 14.2.2, “Scrolling through the Event Log”
Section
14.2.5, “Setting the Event Log Overwrite
Mode”
4
Event dates and times
The dates and times when events occurred are listed here.
You can set the date and time in the system settings by
following the procedures in section 5.5, “Setting the Date and
Time.”
5
Event channels
The channels that events occurred on (A or B) are listed
here.
6
Sync signals
The sync signal statuses (INT or EXT) when events occurred
are listed here.
7
Input formats
The input formats when events occurred are listed here.
NO_SIGNAL
This message appears if there was
no signal.
FORMAT_UNKNOWN This message appears if the input
format could not be determined.
FAN ALARM
This message occurs if the fan was
not operating properly.
8
Event type
The types of events that occurred are listed here.
When the same kind of event occurs successively or when
multiple events occur at the same time, they are treated as a
single event in the event log. You can view all of the events
by saving the event log to USB memory. This is especially
useful when multiple events occur at the same time and you
cannot view all of them on the LV 5330 screen.
The displayed event types are listed below. If the detection of
a particular error has been disabled, its corresponding event
will not be recorded.
TRS TRS
error
LINE
HD-SDI signal line number error
CRC_L
HD-SDI Y signal transmission error
CRC_C HD-SDI
C
B
,C
R
signal transmission error
EDH
SD-SDI signal transmission error
GMUT Gamut
error
CGMUT
Composite gamut error
PRTY
Ancillary data parity error
CHK
Ancillary data checksum error
BCH
Embedded audio transmission error
CRC_WAR
Channel status FORMAT is Consumer
CRC_ERR
Embedded audio CRC error