5. System Settings
5-7
5.4
Configuring the External Interface
To configure the external interface settings, press F•4 INTERFACE in the system menu. A
menu appears for configuring remote control and Ethernet settings.
SYS
→
F•4 INTERFACE
→
Figure 5-4 INTERFACE menu
5.4.1
Selecting the Method for Loading Presets
You can load presets using remote control connector pins /P1 through /P8. There are two
different methods for loading presets. To choose which method to use, follow the procedure
below.
Reference: Section
17.1, “Remote Control Feature”
Procedure
SYSTEM
→
F•4 INTERFACE
→
F•1 REMOTE
Settings
BIT:
/P1 through /P8 are assigned to preset numbers 1 through 8, and you can
load one of eight presets. This is the default setting.
BINARY:
/P5 is set to the MSB, and /P1 is set to the LSB. You can load one of 30
presets by specifying a binary value.