SoundAdvisor Model 831C
Troubleshooting a Measurement Setup File
6-11
6.3 Troubleshooting a Measurement Setup File
Setting the LD Default measurement setup as the Active setup can clear the
meter when it’s doing something unexpected and you don’t know why.
Step 1
On the meter, select
Menu
Setup Manager
.
Step 2
Select
Default
by highlighting it, then pressing
5
. This opens a pop-up menu.
Step 3
Highlight, select
Set to Active
, and confirm your choice if required.
Step 4
Press the
Close
softkey to save the meter settings.