![Larson Davis SoundAdvisor 831C Reference Manual Download Page 134](http://html1.mh-extra.com/html/larson-davis/soundadvisor-831c/soundadvisor-831c_reference-manual_3710133134.webp)
SoundAdvisor Model 831C
Setting Advanced Event Triggers
17-7
Before you begin:
•
This section explains an advanced application of step 4 in
"Setting Up the Event History Measurement" on page 17-1
Complete steps 1–3 before proceeding.
Step 1
On the Event Triggers tab, select
Running Leq
from one of the
Trigger Source
drop-down menus, and enter a
Trigger Level
in dB.
Step 2
Use the right or left softkeys (
*/*
) to navigate to the
SLM
Tab of the Setup
Manager as shown in
Step 3
Enter a
Running Leq
period in seconds. Enter whole numbers only, such as 1–
3600 (1 hour).
FIGURE 17-7
Setup Manager: SLM Tab
In this example, the period is set to be the Leq of the last 1200 seconds of
the measurement run time.
Step 4
Return to Step 5 of
17.1 "Setting Up the Event History Measurement" on page
.
17.4.2 Using 1/1 Octave and 1/3 Octave Event Triggers
Before you begin:
•
This section explains an advanced application of step 4 in
"Setting Up the Event History Measurement" on page 17-1
Complete steps 1–3 before proceeding.
•
Purchase and install firmware option 831C-OB1.
Step 1
On the 831C (or in G4), go to the
Setup Manager
System
Properties
Reference Spectra
tab.