![Larson Davis SoundAdvisor 831C Reference Manual Download Page 132](http://html1.mh-extra.com/html/larson-davis/soundadvisor-831c/soundadvisor-831c_reference-manual_3710133132.webp)
SoundAdvisor Model 831C
Viewing Event History Data on the Events Tab
17-5
FIGURE 17-3
Event History Tab
1.
Green line represents the Trigger Level
2.
Selected event number/number of events or event
records
3.
The 3 events are 3 instances in which the noise
level exceeded the Trigger Level for the Minimum
Duration as indicated in the Setup.
4.
Event Indicator. For details, see
.
FIGURE 17-4
Event Indicator
1.
Ready
When this symbol displays, all the pre-samples are recorded and meter is
waiting for an event.
2.
Trigger
When this symbol displays, the sound level exceeds the Trigger Level.
3.
Valid
When this symbol displays, the sound level exceeds the Trigger Level for the
Minimum Duration.
4.
Continue?
When this symbol displays, the sound level is below the Trigger Level and is in
the Continuation Period until the event continues or the event ends.
17.3.1 Viewing Octave Band and Spectral Events
You can view the resulting events during a measurement in progress on the
Live View
Events
tab on pages 3–6, or after the measurement from the
Measurement
Events
tab on pages 3–4. If you enabled Spectral Mode in the
measurement setup, the Events tab displays multiple pages of data.
1.
2.
4.
3.
1.
2.
4.
3.