![Larson Davis SoundAdvisor 831C Reference Manual Download Page 59](http://html1.mh-extra.com/html/larson-davis/soundadvisor-831c/soundadvisor-831c_reference-manual_3710133059.webp)
SoundAdvisor Model 831C
Creating a Custom Setup File On the Meter
6-3
Step 1
On the 831C, navigate to the Setup Manager File List (first page).
Step 2
Highlight and select your setup file in the list (
8
or
2
+
5
). The name of
your setup displays in the
Settings:
line (see
).
Step 3
Use the right (
//*
) softkey to navigate to the
General
tab as shown in
. These fields are optional.
FIGURE 6-3
Setup Manager: General Tab
1.
When set, this value appears as a filename prefix
for all measurement data files saved using this Setup.
2.
Enter up to 250 characters describing the specif
-
ics of this measurement setup. The description
appears in the Notes section of the downloaded data
file.
Step 4
Use the right (
//*
) softkey to navigate to the
SLM
tab as shown in
.
FIGURE 6-4
Setup Manager: SLM Tab
1.
Select a
Frequency Weighting
,
Detector
, and
Peak
Weighting
. Your choice for the Detector setting com
-
municates to the meter hardware how quickly it needs
to begin measuring the current sound. See also “Detec
-
Appendix C Glossary on page C-1.
2.
If you’re unsure which Integration Method to
choose, select
Linear
, and see the section below called
.
3.
Select the
20 dB Gain
checkbox when measuring
low-level sounds, or when the sound level is close to
the background noise level. For more information, see
the section below called
4.
The
Running Leq
Period
is used when the Event
History option (831C-ELA) is purchased and enabled on
your meter. In this example, the period is the Leq of the
last 1200 seconds of the measurement run time. For
.
1.
2.
1.
3.
4.
2.