55
|
www.lakeshore.com
3.3.8.1 Inspecting the CCR Components
Inspect the CCR compressor for any visible signs of damage caused during shipping,
and for evidence of oil leakage. For the standard water-cooled CCR compressor
configured with the Model CRX-6.5K probe station, the static pressure reading should
be as listed in TABLE 3-6 when the compressor is at room temperature. If any damage
is suspected or the gauge reads less than the specified value, contact your Lake Shore
representative prior to the scheduled installation.
3.3.9 Unpacking the
Options
There are various options that you may have purchased with your probe station.
D
Model TPS-FRG:
the turbo vacuum pumping system option will be contained in two
boxes, one containing the pumping station itself and one containing the gauge
and cables. Remove the gauge and cable components from the box in prepara-
tion for completing the system setup and assembly. Remove the turbo pumping
station from its box. The turbo vacuum pump can be placed on the floor near the
probe station or placed on a cart.
D
PS-TP-KIT:
the turbo pumping kit consists of a 1 meter long NW40 flexible stainless
steel vacuum line in a plastic bag. A separate plastic bag contains the fittings nec-
essary to connect the turbo pumping system to the probe station. Remove the
flexible stainless steel vacuum line and the fittings from the bags in preparation
for completing the system setup and assembly.
D
PS-PLVI:
the pump line vibration isolator will be packed in the crate. The bucket
will typically not be packed in a box, but will be bubble wrapped and shrink
wrapped. The stainless steel vacuum line and associated hardware will be bubble
wrapped and packed in the accessories box.
D
PS-HTSTAGE:
typically, the high temperature stage option will be installed in the
probe station to ensure it is not damaged during shipping.
Compressor
20 ° C (68 ° F)
HC4-E1
1.86 MPa to 1.90 MPa (270 psig to 275 psig)
TABLE 3-6
CCR compressor static pressure